Instrumentation & Control

Measuring flow — orifice plates and the alternatives

Why most flow measurement is really a pressure measurement, what the square root relationship means for turndown, why straight lengths are a measurement requirement rather than a layout preference, and how to choose between orifice, vortex, magnetic, Coriolis and ultrasonic meters.

BeginnerOil & GasPetrochemicalPharmaceutical

Standards referencedISO 5167API MPMSASME MFC-3M

A plant measures flow more often than it measures anything else, and most of those measurements are not really flow measurements at all. They are pressure measurements with some arithmetic on the end.

Understanding that one fact explains most of what follows — including why so many flow readings are quietly wrong.

The differential pressure family

Put a restriction in a pipe. The fluid speeds up to get through it, and as it speeds up its pressure drops. Measure the pressure before and after, and the difference tells you how fast it was going.

The commonest restriction is an orifice plate: a flat disc with a hole, clamped between two flanges. It is cheap, it has no moving parts, it can be pulled out and checked, and it has been the industry default for a century.

Somewhere in the loop the square root has to be taken to turn differential into flow. Modern transmitters do it internally; older systems did it in the control system. Applying it in both places is a classic commissioning error, and it produces a reading that looks plausible at full flow and is badly wrong everywhere else.

Flow measurement principles and where each one fitsSource: ISO 5167 for differential pressure devices; API MPMS for custody transfer
Orifice plateDifferential pressure across a sharp-edged restriction3:1 to 4:1HighClean gas, steam and liquid — the industry defaultLow flow, dirty or erosive fluid, and anywhere the straight lengths cannot be found
VenturiDifferential pressure across a gradual contraction3:1 to 4:1Low — most of it is recoveredLarge lines where pumping cost over plant life outweighs capital costLong, heavy and expensive for the same measurement
Averaging pitotVelocity profile sampled across the pipe3:1Very lowLarge ducts, retrofits, anywhere the line cannot be broken intoMore sensitive to a distorted profile than a full-bore meter
VortexFrequency of vortices shed behind a bluff body10:1 to 20:1ModerateSteam and clean gas or liquid over a wide rangeSignal disappears at low flow; vibration can be read as flow
MagneticVoltage induced as a conductive fluid crosses a magnetic field20:1 or betterNone — full boreWater, slurries, conductive chemicalsHydrocarbons. They are not conductive, so there is no signal at all
CoriolisTwist of a vibrating tube, which gives MASS directly20:1 or betterModerateCustody transfer, varying density, anywhere mass is what is wantedCost, size, and entrained gas in a liquid upsets it
TurbineRotor speed in the stream10:1ModerateClean, low-viscosity liquids; hydrocarbon custody transferIt is a bearing in the process. Particles wear it and viscosity shifts its calibration
Ultrasonic (clamp-on)Transit time of a pulse with and against the flow10:1 or betterNoneTemporary survey, retrofit, no shutdown availableDepends on assumed pipe and fluid properties it cannot see for itself

Turndown is the ratio of the largest to the smallest flow a meter measures usefully. It is the number that most often decides the choice, because a meter that is accurate at design flow and blind at 20 % of it is no use on a plant that spends half its life turned down.

Straight lengths are not negotiable

The published coefficient for an orifice plate assumes the flow arriving at it has a fully developed velocity profile — fastest in the middle, slowest at the wall, symmetrical.

A bend, a tee, a partly open valve or a reducer close upstream leaves the profile lopsided. The plate still produces a differential, the transmitter still produces a number, and the number is wrong by an amount nobody can calculate after the fact.

The plate itself

  • The sharp edge faces upstream. Always. It is what creates the defined contraction.
  • The tab is marked so the orientation can be verified without dismantling anything.
  • A worn or nicked edge reads low, because the contraction is no longer sharp. Erosive service dulls a plate steadily and the drift is gradual enough to go unnoticed.
  • Drain and vent holes sit at the bottom for gas service and the top for liquid, so condensate or gas cannot collect against the plate.

The impulse lines will beat the meter

Everything in the installation topic about impulse lines applies doubly here, because a differential device is measuring a small difference between two large pressures. Anything that affects one leg and not the other appears as flow.

Impulse line arrangements: a gas line rising to a transmitter mounted above the tapping, a liquid line falling to a transmitter below it, and the failure case where liquid collects in a gas leg and adds a false head.
Both legs must see the same temperature and the same fill. A difference between them is indistinguishable from a genuine differential.

Keep the two legs the same length, on the same route, at the same temperature. One leg in the sun and the other in shade produces a differential that changes with the weather — and the resulting fault is reported as an unstable flow transmitter for months.

The meters that are not differential pressure

Vortex. A blunt body in the stream sheds vortices, and the frequency is proportional to velocity. Good turndown, no moving parts, popular on steam. Its weakness is the bottom of the range, where shedding stops and the signal vanishes. Pipe vibration can also be read as flow.

Magnetic. A conductive fluid moving through a magnetic field induces a voltage. Full bore, no pressure loss, unbothered by solids — excellent on water and slurries. Useless on hydrocarbons, which conduct nothing.

Coriolis. A vibrating tube twists in proportion to mass flow. It gives mass directly, so no density figure is needed. That is why it dominates custody transfer. It is expensive, bulky, and upset by gas entrained in a liquid.

Turbine. A rotor spun by the flow. Accurate on clean light liquids, and it is a bearing sitting in the process — particles wear it and viscosity changes its calibration, so it drifts in a way that looks like a process change.

Variable area — the rotameter. A float rides in a tapered glass or metal tube; the greater the flow, the higher it sits, and you read the scale beside it. It needs no power and no signal, which is why purge and seal flows all over a plant are set with one.

It must be mounted vertically, because the measurement is a balance between the drag of the flow pushing up and the weight of the float pulling down. Fit one horizontally and the float simply rests against the tube wall — the reading has no meaning at all.

Ultrasonic clamp-on. Times a pulse with and against the flow. Nothing enters the pipe and nothing has to be shut down, which makes it ideal for a survey. The price is that it is working from assumed pipe wall and fluid properties it cannot actually see.

Density, and why gas flow needs compensating

An orifice plate responds to the kinetic energy of the stream, which depends on density as well as velocity. On a liquid, density barely moves and can be treated as fixed. On a gas it moves with pressure and temperature, and it moves a lot.

Without compensation, the indicated flow changes when the line pressure changes although the actual flow has not moved. Compensated measurement takes pressure and temperature signals and corrects continuously — which is why a gas flow measurement is often three instruments, not one.

What to take away

  • Most flow measurement is a differential pressure measurement with a square root on it.
  • Flow goes as the square root of differential. That relationship is what limits an orifice meter to roughly 3:1 or 4:1 turndown.
  • Straight lengths are a measurement requirement. Lose them and the error cannot be calculated afterwards — only estimated.
  • The sharp edge faces upstream; the tab exists so you can check without pulling the plate.
  • Both impulse legs must be the same length, route and temperature. A difference between them is read as flow.
  • Magmeters do not work on hydrocarbons. Coriolis gives mass directly. Vortex loses the signal at low flow. Turbine is a bearing in your process.
  • Gas flow needs pressure and temperature compensation, or the reading moves when the pressure does.
  • A flat trend on a live process usually means a blocked impulse line, not stability.

Check your understanding

6 questions. Nothing is recorded — this is just for you.

1An orifice plate produces four times the differential pressure it did an hour ago. What has happened to the flow?
2Why does an orifice plate need a defined straight length of pipe upstream?
3A magnetic flowmeter is proposed for a naphtha line. What is the problem?
4What does a Coriolis meter measure that the others do not?
5An orifice plate is found installed with the sharp edge facing downstream. What is the consequence?
6Why does a gas flow measured by orifice plate need pressure and temperature compensation?

#instrumentation#flow#orifice#measurement#transmitter